Pressure Gauge Calibration in Oxnard, CA

QSI calibrates pressure gauges, pressure transducers, vacuum gauges, differential pressure instruments, and barometers at our ISO/IEC 17025 accredited calibration laboratory in Oxnard, California. Our capabilities range from vacuum up to 12,000 psig, with deadweight testers as reference standards from 1 psig up. Ship your instruments to our lab from anywhere in the U.S., or schedule on-site calibration in Ventura County, Santa Barbara County, the Santa Clarita Valley, and Los Angeles.

Pressure Equipment We Calibrate

  • Pressure Gauges

  • Pressure Transducers

  • Vacuum Gauges

  • Differential Pressure

  • Barometers

Don't see your equipment listed? Contact us for a custom quote.

Why Deadweight Testers Matter

A deadweight tester generates pressure by placing precisely known masses on a piston of precisely known area. Because the pressure comes directly from mass and area, not from another electronic sensor, deadweight testers are among the most accurate and stable pressure references available. When your gauges are calibrated against a deadweight tester, you're comparing them to a reference that's far more accurate than the instruments being tested.

Pressure Gauge Calibration

Pressure gauges drift for predictable reasons. Pressure spikes and pulsation fatigue the mechanism inside dial gauges. Overpressure, even briefly, can permanently shift readings. Vibration wears moving parts, and temperature swings affect both mechanical and digital sensors. A gauge with a needle that still moves smoothly can easily be reading high or low.

Calibration applies known pressures at multiple points across the gauge's range and documents how accurately it reads at each one. Checking points on the way up and on the way down also reveals hysteresis, where a gauge reads differently depending on whether pressure is rising or falling.

Vacuum, Differential Pressure, and Barometric Instruments

Vacuum gauges, differential pressure instruments, and barometers each measure pressure relative to a different reference: a vacuum, a second pressure, or the atmosphere. Differential pressure instruments are especially common in cleanrooms and controlled environments, where they confirm that rooms hold the pressure differences that keep contamination out. Each type needs calibration against the right kind of reference to ensure its readings mean what you think they mean.

Who Relies on Pressure Calibration


Pressure measurement runs through nearly every industry in QSI's service area.

Aerospace and defense: hydraulic and pneumatic systems, test stands, and ground support equipment

Pharmaceutical and biotech: differential pressure monitoring for cleanrooms and controlled environments

Medical device manufacturing: leak testing and process gauges under ISO 13485 quality systems

Energy and utilities: process gauges on oil, gas, and water systems

General manufacturing: hydraulic presses, compressed air systems, and process equipment

Pressure Gauge Calibration

Frequently Asked Questions